Overview
- Contribute to measurement safety in high voltage environments
- High-quality floating measurement is possible with excellent common mode removal performance
- Extremely effective for noise tolerant measurement
- Compatible with a wide range of applications such as lightning surges and charging tests
- Remote point measurements (switches, transport equipment, etc.)
- Failure cause analysis when an abnormal location occurs in a remote location
(Up to four isolation units can be installed separately)
Appearance
Isolation Probe
Measurement application
Connection image
Isolation probe SE-6000 / SE-6011 specifications
Item | Specifications |
---|---|
Number of channels | 1 to 4 CH (up to 4 SE-6011s can be connected per SE-6000) |
Frequency bandwidth (at the 50Ω termination) |
30MHz -3dB |
Rise time (at the 50Ω termination) |
13.3ns (typical) |
Receiving side Output impedance |
50Ω±1.5% |
Bandwidth limit | 10MHz ±3MHz, 1MHz ±0.3MHz, 100kHz ±30kHz Attenuation characteristic:-6 dB/oct |
Probe sense | ■ Auto search 1:1, 10:1, 100:1, 1000:1
■ Manual setting 1:1, 5:1, 10:1, 20:1, 50:1, 100:1, 200:1, 500:1, 1,000:1, 2,000:1 |
CMRR | 100 dB (at 80Hz) |
Frequency flatness (at the 50Ω termination) |
DC~5MHz:±0.3dB 5MHz~10MHz:±1dB |
Input impedance | 1MΩ(±1%) // 20pF(±2pF) |
Input range (Full scale) |
■ When an output range of ±1V (50Ω) / ±2V (1MΩ) is selected ±50mV, ±100mV, ±200mV, ±500mV, ±1V, ±2V, ±5V, ±10V, ±20V, ±50V ■ When an output range of ±0.8V (50Ω) / ±1.6V (1MΩ) is selected ±40mV, ±80mV, ±200mV, ±400mV, ±800mV, ±2V, ±4V, ±8V, ±20V, ±40V The input voltage range is the input range setting value - offset setting value |
Input coupling | DC, AC, GND |
AC coupling lower limit frequency (-3dB) |
<10Hz |
Output voltage | ■ When an output range of ±1V (50Ω) / ±2V (1MΩ) is selected
50Ω termination ±1V 1MΩ termination ±2V ■ When an output range of ±0.8V (50Ω) / ±1.6V (1MΩ) is selected 50Ω termination ±0.8V 1MΩ termination ±1.6V |
Noise (RMS) | 0.05%FS |
DC gain accuracy | ■ At 1MΩ termination After self-calibration at operating temperature ±0.3% At operating temperature ±1% ■ At 50Ω termination Add the following to the value at 1MΩ termination. (50Ω termination resistance error [%] - SE6000 output impedance error [%]) / 2 approximation formula) |
Offset voltage range | Up to ±500mV range:±1V ±800mV to ±5V range:±10V ±8V to ± 50V range:±100V |
Offset accuracy | ■ After self-calibration at operating temperature ±0.5% (set value) ±0.3% FS ±0.05mV (input conversion value) ■ Operating temperature ±0.5% (set value) ±5% FS ±1mV (input conversion value) |
Nonlinearity | ±0.05%FS |
Maximum input voltage | Between BNC center contact and BNC shell:400Vpeak Between BNC shell (housing) and earth:1kVpeak |
Propagation delay | 1.089µs±0.015µs+5ns/m 5ns/m is the reference value for optical fiber cable delay. |
Between CH skew | Within 20ns (excluding cable delay difference) |
Calibration output | Square wave / 1kHz ±0.5% Output amplitude / 3V ±2% |
Self-calibration | Self-calibrate amplitude and offset |
Preheating time | 30 minutes |
Auto range | Automatic adjustment of offset and range |
Save/recall | Save destination:USB memory, internal memory (up to 10) / target data:Setup |
Isolation amplifier SE-6000 (receiving side) specifications
Item | Specifications |
---|---|
Display | 4.3“ color LCD |
Interface | USB:USB2.0 host (for save/recall)
LAN:100Base-TX (for remote control) |
Power supply A | AC 100V~240V (50/60Hz) |
Power consumption | Up to 30VA |
Operating temperature | 0ºC~+40ºC |
Performance warranty temperature | +10ºC~+35ºC |
Operating humidity | 5% to 80% RH (no condensation) |
Storage temperature | -20ºC~+60ºC |
Storage humidity | 5% to 80% RH (no condensation) |
Altitude | 2,000m or less |
External dimensions | Approx. 226 W x 100 H x 366 D mm (excluding protrusions) |
Main unit weight | Approx. 2.5kg |
Accessories | User’s guide, instruction manual (CD), power cord, cord strap 1 each |
Isolation unit SE-6011 (transmitting side) specifications
Item | Specifications |
---|---|
Battery (Optional) |
A maximum of 2 pieces can be installed * Please be sure to purchase. |
Operating time | New product, fully charged, at 25°C 1 battery:15 hours 2 batteries:30 hours |
Operating temperature | -15ºC~+50ºC |
Operating humidity | 5% to 80% RH (no condensation) |
Storage temperature | -20ºC~+60ºC |
Storage humidity | 5% to 80% RH (no condensation) |
Altitude | 2,000m |
External dimensions | Approx. 90W x 203H x 298D mm |
Main unit weight | Approx. 2.4kg (excluding battery) |
Optical interface
Item | Specifications |
---|---|
Light source semiconductor laser | Class 1 |
Optical fiber | Single mode |
Optical connector | SE-6011:Duplex LC connector × 1 SE-6000:Duplex LC connector x 4 |
Configuration
Product Name | Model Name | |
---|---|---|
Isolation amplifier (receiving side) | SE-6000 | |
Isolation unit (transmitting side) | SE-6011 | |
Battery charger (2 pcs can be charged at the same time) * Necessary when charging the “dedicated Li-ion battery” | CH5050A | |
Battery (for SE-6011, up to 2 batteries can be installed) | NH2054 | |
Optical fiber cable | 3m | SE-605 |
10m | SE-606 | |
50m | SE-607 (Made-to-order) | |
200m | Please contact us. |