Universal Counter SC-7205A Options

GPIB Interface SC-701 (for SC-7205A)

Factory Option. Please order at the same time as the main unit.
The universal counters SC-7207H, SC-7206H, SC-7205H, and SC-7205A have a built-in SC-701 (SC-7207H, SC-7206H are standard equipment), which enables remote setting and external transmission of measurement result data related to measurement reset, measurement function, time base, calculation, etc. by an external GP-IB controller.

  • Electrical and mechanical specifications are in accordance with IEEE 488.1.
  • It has the following interface functions for GPIB.
    SH1, AH1, T5, L4, SR1, RL1, PP0, DC1, DT1, C0
  • It has a REMOTE/LOCAL status LED indicator.
  • Fully remote control supported (except for some functions such as on/off of the power switch and remote related settings.)
Equipped model
SC-7207H Standard equipment
SC-7206H
SC-7205A
SC-7205H
Option

Address settings

The following settings are possible through the SYSTEM menu of UTILITY.

  • The 31 types of device addresses from 0 to 30 can be selected at will.
  • One of CR+LF or LF can be selected as the delimiter.

Digital I/O SC-702 (for SC-7207H, SC-7206H, SC-7205H, SC-7205A)

Factory Option. Please order at the same time as the main unit.
By integrating the SC-702 (digital I/O), the universal counter SC-7205A can control the start of measurement and output the comparator operation result (open collector). By connecting lamps, etc. to the exterior, you can easily visualize parts sorting and inspection results.

Timing chart

Control signal

/TRIG C-MOS input Start the measurement with an L pulse of 10ms or more during the /BUSY = Z period.
The /TRIG signal during the /BUSY = L period is invalid.
/BUSY open collector output Measuring L (output transistor on), non-measuring Z (output transistor off)
/HI, /LO, /GO
Open collector output
One place is L (output transistor on) according to the comparator operation result

Otherwise, Z (output transistor off)

Output terminal maximum setting

  • Withstand voltageDC50V
  • Withstand current DC150mA
  • Frequency responseDC~1kHz

Input terminal maximum rating

  • Withstand voltageDC5V
  • Frequency responseDC~1kHz

High stability reference oscillator SC-703A

Factory Option. Please order at the same time as the main unit.

Item Performance
Oscillation frequency ±0.05ppm
+25°C as reference 0°C to +40°C range
Rise time ±0.05ppm
10 minutes after power on based on the frequency 1 hour after power on
Change over time (/day) ±0.02ppm
Value after 72 hours based on 48 hours after power is turned on
Change over time (/month) ±0.1ppm
Value after one year based on 10 days after power is turned on

Other options

Model Overview Appearance
SC-525 USB-RS converter. By adopting a cross cable and using the RS connector as a female connector, it is possible to directly connect the USB port and the measurement equipment. You can use an external controller (usually using a PC) to perform the same operations as when you manually control using button operation on the main unit. Not only operations, but also measurement result data collection is possible (interface specifications are the same as the standard RS-232).
BB-120CA Coaxial cable(BNC - BNC) 120cm type
BB-150CA Coaxial cable(BNC - BNC) 150cm type
Input coaxial cable B (BNC - crocodile port [black, red])